Abstract: A novel artifact zirconia-dolerite ball plate is developed for interim check of coordinate measuring machine. Artifact contains 52 balls and distances between balls are used for analysis of measuring volume of CMM. Statistical performance analysis is explained in this paper. Stability of Ball plate is explored through cross correlation. The resultant corroleogram depicts that the ball plate can be used for 11 months without recalibration or rechecking for dimensional stability.
Keywords: Artifact, Ball Plate, Run Test, Cross Correlation